'''Device name:''' |
'''Device name: Gas-Pycnometer AccuPyc II 1340''' |
||
'''Operator/Responsible for this device:''' |
'''Operator/Responsible for this device: [[User:Frieder.Koch | Frieder Koch]]''' |
||
'''Position:''' |
'''Position: clean room, R 307 room 132c''' |
||
'''Device designation:''' |
'''Device designation:''' |
||
'''Specifiaction number:''' |
'''Specifiaction number: SP-281071''' |
||
<!-- Designation as per the list of fabrication and measurement devices list, (Fertigungs- und Prüfmittelliste) , e.g. (MI006) --> |
<!-- Designation as per the list of fabrication and measurement devices list, (Fertigungs- und Prüfmittelliste) , e.g. (MI006) --> |
||
== Short Description == |
== Short Description == |
||
The AccuPyc is a device for measuring the volume of a sample at high precision. It is based on Boyle's law and measures the amount of gas displaced by the sample. |
|||
== Alternatives == |
== Alternatives == |
||
| Line 27: | Line 28: | ||
== Device Specifications == |
== Device Specifications == |
||
(Additional) equipment and functions, e.g. process gasses, wavelengths, resolution, smallest achievable feature size |
|||
(Zusätzliche) Ausstattung und Funktionen, z.B. Prozessgase, Wellenlängen, Auflösung, Strukturdimensionen |
|||
The device is currently operated with Nitrogen as analysis gas. |
|||
== Established Processes == |
|||
Which processes have been compelted successfully with this device, what materials have been structured, testet or processed? |
|||
The available sample chambers differ in size : |
|||
Please don't enter any process parameters and sensible details into the wiki, but add the path to J: network drive where the detaisl are described or link a user who knows about the process. |
|||
*100 cm³ |
|||
*35 cm³ |
|||
*10 cm³ |
|||
The analysis temperature can be set freely between room temperature and 90 °C. |
|||
The accuracy of the device is 0,03 % of the sample chamber size plus 0,03 % of the sample volume. |
|||
== Restrictions == |
== Restrictions == |
||
What can not and/or should not be done with the device, i.e. are there materials that could impair the performance of the device if they are structured with, e.g. nickel inside the RIE? Are there other precautions that have to be taken for sample preparation? |
|||
The device is not suited for liquids. The size of the sample is limited by the cylindric sample chamber (diameter 4,6 cm, height 6,17 cm). |
|||
== Instructed Users == |
== Instructed Users == |
||
Persons who have recieved an instruction and thus are allowed to operate the device. |
|||
[[User:Frieder.Koch | Koch, Frieder]] |
|||
Please create a list and link them to the user profiles: |
|||
* [[Benutzer:Vorname.Nachname | Nachname, Vorname]] |
|||
Latest revision as of 19:48, 20 November 2016
Device name: Gas-Pycnometer AccuPyc II 1340
Operator/Responsible for this device: Frieder Koch
Position: clean room, R 307 room 132c
Device designation:
Specifiaction number: SP-281071
Short Description
The AccuPyc is a device for measuring the volume of a sample at high precision. It is based on Boyle's law and measures the amount of gas displaced by the sample.
Alternatives
Are there other devices at IMT with the same or a similar application? Please name them and their differences or similarities, and link them here. YOu can find the correct link name in the list of measurement devices or in the list of fabrication devices.
Device Specifications
The device is currently operated with Nitrogen as analysis gas.
The available sample chambers differ in size :
- 100 cm³
- 35 cm³
- 10 cm³
The analysis temperature can be set freely between room temperature and 90 °C.
The accuracy of the device is 0,03 % of the sample chamber size plus 0,03 % of the sample volume.
Restrictions
The device is not suited for liquids. The size of the sample is limited by the cylindric sample chamber (diameter 4,6 cm, height 6,17 cm).