Oberflächenmessgerät TENCOR P 2
From IMT-Wiki
Jump to navigationJump to search
The purpose of these instructions is to give you a quick overview on how to work with the devices. To use the device please contact the responsible person to receive a proper introduction.
Device location: Resistlabor
Phone number: 22723
Lab responsible: Birgit Hübner
Principal users: Heike Fornasier, Margarita Zakharova, Abrar Faisal, etc
Access: the device is located in the end of Resist Lab, near the optical microscope.
Short Description
The Tencor P2 Long Scan Profiler is a computerized, high-sensitive surface profiler that measures step height and roughness in a variety of applications.
Start-up Procedure
- Enable p2. Power should be on all the time! (If power is off: Press the Power Button: Turn AC power by pressing the Red Power switch to the on the position on Auxiliary Panel. The Operating system will boot up automatically.
- Press “Enter” Key after Logo appears on the screen.
Mounting the Sample
- Open the Measurement Area Door. Stage should be in front, measurement head including stylus up.(If not: Press the LOAD button on the Built-in Keyboard or the F7/LOAD-UNLOAD button on the Remote Keyboard.)
- Place the sample on the stage.
- Press the LOAD button (stage moves back).
- Close the Measurement Area Door.
Lowering the Measurement Head Unit
- Press the Zq button on the Built-in keyboard or press F5/ Zq button of the Remote Keyboard.
- Press the [ ¯ ] button on both keyboards for Low speed head motion.
- Press the SEL + [ ¯ ] button on the Built-in Keyboard and Space bar + [ ¯ ] button on the Remote Keyboard for High Speed Head motion. CAUTION: Do not lower the head onto the sample in high speed mode!
- Lower Measurement Head unit till the stylus makes contact with surface of sample and release [ ¯ ] button.
Optical Zoom
Optical Zoom can be enable when sample has been loaded and measurement head is lowered and focus on sample.
- Press the Zoom button on the Built-in Keyboard or the F11/ZOOM button on the Remote Keyboard.
- Press the [ ¯ ] button to decrease magnification.
- Press the [ ] button to increase magnification.
- Press the XY or Zq button for motion control.
Selecting a Scan Recipe
- Press the MENU button on the Built-in Keyboard or the ESC/MENU button on the Remote Keyboard.
- Select the Recipe Menu.
- Select the Catalog. Press Enter.
- Select the desired Recipe. Press Enter.
Unloading of Sample
- Press LOAD, stylus moves up and stage moves forward.
- Open the Measurement Area Door.
- Remove sample from stage.
- Close the Measurement Area Door